EasyManua.ls Logo

Sotera VISI - Managing Alarm;Alert Annunciations

Default Icon
242 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Physiological Alarms (Alarms) / Technical Alarms (Alerts) Summary
84
6.2.2 Managing Alarm/Alert Annunciations
Annunciation: Refers to the audible and visual display of alarms/alerts.
Note: When multiple alarms and alerts occur simultaneously, the message text will display
messages associated with only the highest alarm severity. The vital signs measurements will
display all existing alarms, regardless of their severity.
Action Action Button Duration Allow Where Effects
Pause Alarms/
Alerts
N/A 2 Minutes May only be paused
from the ViSi Mobile
Monitor.
Annunciation disabled at
both the ViSi Monitor and
Remote Viewer.
Resume Alarms/
Alerts
N/A ViSi Mobile Monitor and
Remote Viewer.
Annunciation resumed at
both the ViSi Mobile
Monitor and Remote
Viewer.
Turn Alarms/
Alerts Off
N/A Indefinitely May only be turned off
from the ViSi Mobile
Monitor.
Annunciation disabled at
both the ViSi Monitor and
Remote Viewer.
Turn Alarms/
Alerts On
N/A ViSi Mobile Monitor and
Remote Viewer.
Annunciation turned on at
both the ViSi Mobile
Monitor and Remote
Viewer.

Table of Contents