Cautions
26
The ViSi Mobile Monitoring System may not perform to specification if stored or shipped
outside the specified temperature range.
The ViSi Mobile Monitor may be temporarily interrupted by UHF RFID Systems (860-
960MHz).
When using a ViSi Power Pack equipped with a mount, ensure the clamp is properly
secured to the bedside or IV Pole to avoid damage from being dropped.
Avoid from putting the ViSi Power Pack directly below an IV bag.
Avoid putting anything other than the ViSi Power Pack into the cradle.
Route the ViSi Power Pack cable away from other medical equipment in its vicinity.
Cleaning / Disinfecting
Do not clean the ViSi Mobile Monitor, the Cuff Module, or the Power Pack while it is
plugged into the ViSi Mobile Charger.
Do not apply liquid to the ViSi Mobile Cuff Module or the Power Pack. To clean, use a
damp cloth.
Ensure the sensor connector contacts are thoroughly dried to prevent possible
malfunction.
Thumb sensors which are saturated with liquid should be allowed to air dry thoroughly
before re-use.
Do not use bleach, abrasive cleaning agents or organic solvents on any of the ViSi
Mobile Monitoring System components.
Use only recommended cleaning/disinfecting agents to prevent damage to the device
and components. See page 107
Do not autoclave the ViSi Mobile Monitor, its components, or accessories.
Do not use excessive amounts of liquid when cleaning the ViSi Mobile Chest Sensor or
the ViSi Mobile Thumb Sensor.
Cautions
After patient use, the disposables from the ViSi Disposable Kit may contain bio-hazard
materials. Handle and dispose of these items according to your facility’s policies.
When the ViSi Mobile Cuff Module is connected to the other ViSi Mobile Components,
the entire system has an ingress protection rating of IPX0.