Clinical Configurations
78
5.5.2 Units of Measure
5.5.3 Monitor Timeout
The Clinical screen on the Wrist Monitor will timeout after a specified period of inactivity.
a. The Wrist Monitor screen can be set to timeout at different intervals and the user will be
required to authenticate before viewing the screen again.
b. When “Off” has been selected, the screen will never timeout; a user will not be required to
authenticate to view the screen again.
5.5.4 Clinical Authentication
The Clinical Authentication option determines if a user must authenticate using a 4-digit PIN before
accessing the Monitor features.
a. When “On” has been selected, the user will be prompted to authenticate using a 4-digit PIN
before being able to access the clinical screens.
Configuration Options Option Values Level Updateable
Blood Pressure mmHg (default)
kPa
Facility: yes
Care Unit: no
Temperature
Fahrenheit (
o
F) (default)
Centigrade (
o
C)
Facility: yes
Care Unit: no
Date Display Format Mon DD, YYYY(default)
DD Mon YYYY
MM/DD/YYYY
DD/MM/YYYY
Facility: yes
Care Unit: no
Time Display Format 12 hr (default)
24 hr
Facility: yes
Care Unit: no
ECG Line Noise Filter 50 Hz
60Hz (default)
Facility: yes
Care Unit: no
Configuration Options Option Values Level Updateable
Monitor Timeout 15 Seconds
30 Seconds
1 Minute
3 Minutes (default)
Off
Facility: yes
Care Unit: no
Monitor: no
Configuration Options Option Values Level Updateable
Clinical Authentication On (default)
Off
Facility: yes
Care Unit: yes
Monitor: yes