Appendix A - Alarm Summary
Patient Alarms
Life Threatening Alarms
Note: When the Cuff Module is in the Charger, the LED on the front of the Cuff Module will
display red.
Display Message Symbol(s) Alarm Summary Cause
CRITICAL LOW HR Critical Low Heart Rate
Note: Patient’s heart rate is less than 18
BPM.
MONITOR TOO HOT Monitor - Battery Over-
Temperature Failure
Battery in the Monitor has exceeded
a safe temperature.
The Chest Module has exceeded a
safe temperature.
Return the Monitor and the Chest
Sensor to Sotera Wireless, Inc.
CUFF BATTERY
TEMP
Cuff Module - Battery
Over-Temperature
Failure
Battery in the Cuff Module has exceeded
a safe temperature.
Return the Cuff Module to Sotera
Wireless, Inc..
When the “Monitor Too Hot” alarm is in progress, the ViSi Mobile Monitor
and Chest Sensor should be removed from the patient immediately. Leaving
them on the patient for an extended period of time may lead to a skin burn.
When the “Cuff Battery Temp” alarm is in progress, the ViSi Mobile Cuff
Module should be removed from the patient immediately. Leaving it on the
patient for an extended period of time may lead to a skin burn.