Applying ViSi Mobile Monitoring System / Initiate
75
6. Plug the Cuff Module into any of the open ports on the flat end of the Monitor with the connector
contacts facing upwards.
The NIBP Start button will be enabled, prompting you to start a Cuff inflation. See Taking a NIBP
Measurement on page 124 for instructions on how to take a NIBP measurement.
Avoid applying the ViSi Mobile Disposable Cuff over a wound as this can
cause further injury.
Avoid applying the ViSi Mobile Disposable Cuff on any limb where
intravascular access or therapy, or an arterio-venous (A-V) shunt, is present
because of temporary interference to blood flow which could result in injury
to the patient.
Take care in the application of the ViSi Mobile Disposable Cuff when
applying the Cuff to an arm on the same side of a mastectomy. Recommend
using the ViSi Mobile Monitoring System on the opposite arm.
ViSi Mobile blood pressure measurements (NIBP and cNIBP) have not been
clinically evaluated in the presence of atrial or ventricular arrhythmias. Use
alternative BP methods if these arrhythmias are present.
Inflate the ViSi Mobile Disposable Cuff only after proper application to the
patient’s limb.
Insert with connector
contacts facing upwards