Applying ViSi Mobile Monitoring System / Initiate
69
5.3.4 Applying the ViSi Mobile Chest Sensor and ECG Electrodes
Skin Preparation
Skin preparation and ECG electrode placement directly impact the quality of the ECG signal and HR
determinations. Follow your facility’s protocol for skin preparation.
Note: To avoid skin irritation, avoid areas that appear damaged; remove ECG electrodes if the
patient complains of pain/itching; replace ECG electrodes per the electrode manufacturer’s
instructions and place on different sites.
Note: Only use snap-on type electrodes.
Use all of the same type of high quality ECG electrodes on the patient.
Mixing ECG electrode types can adversely affect ECG monitoring.
To ensure patient safety, use only components and accessories
recommended or supplied by Sotera Wireless, Inc. Accessories must
always be used in accordance with your facility’s policies and the
manufacturer’s recommendations.
Only use the ViSi Mobile Chest Sensor CableChest Sensor
provided by Sotera Wireless, Inc. for the ViSi Mobile Monitoring
System. The Chest Sensor CableChest Sensor is designed to
provide defibrillation protection as indicated in the Specifications
section of this manual. ViSi Mobile is designed to be compatible
with the use of external defibrillators.
To ensure patient safety, the conductive parts of the ECG electrodes,
including connectors and other patient-applied components, should not
contact other conductive parts, or earth ground, at any time.