Safety
28 SPECTRO xSORT05.09.2014
2.7.4 Example 4:
Measurement conditions:
Use of the radiation protection shield
The sample consists of a thin plastic material (housing, toy, etc.)
Measuring point 1 2 3
Distance
5 cm
(2.0 inch)
4
Distance
50 cm
(19.7 inch)
5
Distance
200 cm
(79 inch)
Method for analyzing
metallic samples
<1 µSv/h
15 µSv/h
20 µSv/h 4 µSv/h <1 µSv/h
Maximum parameters <1 µSv/h
30 µSv/h
80 µSv/h 8 µSv/h <1 µSv/h
Under the measurement conditions that are stated above, we recommend a
minimum distance of 2.0 m (79") on the side that faces away from the radiation
protection shield.