26 SPECTRO xSORT — 05.09.2014
2.7.2 Example 2:
Measurement conditions:
Use of the radiation protection shield.
The sample thickness prevents the protective sleeve from touching the
radiation protection shield.
Measuring point 1 2 3
15 cm
(5.9")
distance
4
80 cm
(31.5")
distance
5
200 cm
(79")
distance
Method for analyzing
metallic samples
<1 µSv/h <1 µSv/h 30 µSv/h 2 µSv/h <1 µSv/h
Maximum parameters <1 µSv/h 4 µSv/h 70 µSv/h 5 µSv/h <1 µSv/h
Under the measurement conditions that are stated above, we recommend a
minimum distance of 2.0 m (79") on the side that faces away from the radiation
protection shield.