— 05.09.2014 —
2.7.3 Example 3:
Measurement conditions:
Without using the radiation protection plate / the radiation protection shield
The protective sleeve encloses the plastic sample completely (measured
value a)/incompletely (measured value b)
Measuring point 1 Measured value a Measured value b
Method for analyzing
metallic samples
< 1 µSv/h 1 µSv/h
Maximum parameters
< 1 µSv/h 25 µSv/h
These values underline why - for the protection against radiation - the docking
station or the radiation protection shield should always be used for non-metallic
samples.