00732467 Rev. D Page 79 of 110
Table 12-1 (Cont’d)
Text On Screen
Alarm Priority
Action
WARNING:
External Flash Failure
Medium Priority
Finish procedure, then Contact STERIS Customer
Service for equipment service.
WARNING:
Internal Flash Failure
Medium Priority
Finish procedure, then Contact STERIS Customer
Service for equipment service.
WARNING:
Irrigation pump failure.
Medium Priority
Contact STERIS Customer Service.
FAULT:
Disconnect footswitch.
Turn generator OFF.
Wait 3 seconds.
Turn back ON.
High Priority
Disconnect footswitch. Turn generator OFF.
Wait 3 seconds. Power unit ON.
FAULT:
FPGA Communications
Failure. Contact Technical
Services.
High Priority
Contact STERIS Customer Service for equipment
service.
FAULT:
FPGA argon input pressure
failure.
Contact Technical Services.
High Priority
Remove and re-install the Argon Gas Bottle. See
section 9.4.2 System Installation and Setup:
Installing Argon Gas Bottle.
Cycle unit power. If POST is successful, go to Argon
Method and push PURGE. If the fault does not
reappear, the gi4000 is OK to use. If POST is not
successful or if purging triggers the fault screen,
contact STERIS Customer Service for equipment
service.
FAULT:
FPGA argon output
pressure failure.
Contact Technical Services.
High Priority
Check that ArConnect® argon probe connector
and/or the argon coagulation probe is not kinked,
pinched, or crushed.
Check that argon coagulation probe tip is not
blocked. If it is blocked, replace instrument.
Cycle unit power. If the fault does not reappear, the
gi4000 is OK. If POST is not successful or if purge
triggers the fault screen, then Contact STERIS
Customer Service for equipment service.
FAULT:
FPGA argon flow level
failure. Contact Technical
Services.
High Priority
Contact STERIS Customer Service.