00732467 Rev. D Page 80 of 110
Table 12-1 (Cont’d)
FAULT:
FPGA argon bottle high
pressure. Contact
Technical Services.
Contact STERIS Customer Service.
FAULT:
FPGA high voltage
relay check. Contact
Technical Services.
Contact STERIS Customer Service.
FAULT:
Temperature failure.
Turn off and
let cool down.
Power OFF. Let unit cool down before turning unit
ON.
FAULT:
Touchscreen circuit failure.
Contact Technical Services.
Do not touch the Touchscreen during POST. Power
OFF/ON. If the fault does not reappear, the gi4000
is OK. If POST is not successful or if the fault
screen returns, then Contact STERIS Customer
Service.
FAULT:
SLCD macro file
version failure. Contact
Technical Services
Contact STERIS Customer Service.
FAULT:
SLCD communications
failure. Contact Technical
Services.
Contact STERIS Customer Service.
FAULT:
Disconnect Stuck
Footswitch
Footswitch may be sticky or dirty. Soak and clean.
Do NOT submerse Footswitch connector.
FAULT:
Footswitch Circuit Failure
Contact Technical Services
Contact STERIS Customer Service.
FAULT:
Pad circuit failure.
Contact Technical Services.
Contact STERIS Customer Service.
FAULT:
Relay failure.
Contact Technical Services.
Contact STERIS Customer Service.
FAULT:
RF output failure.
Contact Technical Services.
Power OFF/ON. If the fault returns, contact STERIS
Customer Service.