70 6506-209-001 REV F www.stryker.com
Cleaning
Cleaning the battery (Continued)
To clean the SMRT Pak:
1. Remove the SMRT Pak from the Power-PRO cot or SMRT charger.
2. Inspect the SMRT Pak housing and terminal area for any cracks or damage.
3. Clean the SMRT Pak with a hospital grade disinfectant with a pH range of 6.5 to 10.5.
4. Rinse the SMRT Pak thoroughly with clean water to remove any cleaning chemical or residue. Position the SMRT
Pak to avoid water from pooling near the terminals.
5. Dry the SMRT Pak thoroughly before insert the SMRT Pak into a Power-PRO cot or SMRT charger.