72 6506-209-001 REV F www.stryker.com
Preventive maintenance
Regular inspection and adjustments (Continued)
Item Inspect
Head section
Safety bar operates. Pull toward the head section to make
sure that the safety bar swings and rotates freely and
pulls back to the home position
Restraint Restraints function with no excessive wear (such as a
bent or broken receiver or latch plate or torn or frayed
webbing)
Battery
SMRT Pak housing and terminal area for cracks or
damage before first and every use
Charger SMRT charger and parts for cuts in the cord, bent pins or
contacts, or cracks in the housing before first and every
use
Every three months or six hours
Inspect these items every three months or six hours, whichever comes first.
Item Inspect
Cylinder All fasteners are secure
No hydraulic fluid (red) leaks
Loose fittings - tighten, if needed
Hydraulics
Motor mount fasteners are secure
No hydraulic fluid leaks
No leaks from reservoir
Cables and wires No damage or pinching of wiring harness, cable, or lines
No damaged connectors
Manual backup release handle Base extends and retracts when you pull the manual
backup release handle
Cot does not lower when you pull the manual backup
release handle with 100 lb (45 kg) or more on the cot
Litter All fasteners are secure
Backrest cylinder operates
Adjust pneumatic cylinder for full range of motion, if
required
Base All fasteners are secure
X-frame X-frame expands and retracts
Kickstand (optional)
Retracts fully to the transport position
Bolts are tightened