Preventative Maintenance
10
REMOVAL OF IODINE COMPOUNDS
This solution may be used to remove iodine stains from mattress cover and foam surfaces.
1. Use a solution of 12 tablespoons Sodium Thiosulfate in a pint of warm water to clean the stained area.
Clean as soon as possible after staining occurs. If stains are not immediately removed, allow solution to
soak or stand on the surface.
CHECKLIST (PERFORM A MINIMUM OF TWICE A YEAR)
All fasteners secure (reference all assembly prints)
All casters lock with brake pedal engaged
Brake Not Set LED (on foot board) blinks when brakes are not engaged
Locking steer caster engages and disengages properly
Siderails move, latch and stow properly
CPR release working properly
Foot prop intact and working properly
I.V. pole working properly
Foley bag hooks intact
Optional chart rack intact and working properly
Optional CPR board not cracked or damaged and stores properly
No cracks or splits in head and foot boards
No rips or cracks in mattress cover
All functions on head end siderails working properly (including LEDs)
All functions on footboard working properly (including LEDs)
Scale and Bed Exit system calibrated properly
Motion Interrupt switches working properly
Optional night light working properly
Power cord not frayed
No cables worn or pinched
All electrical connections tight
All grounds secure to the frame
Ground impedance not more than 100 milliohms
Current leakage not more than 100 microamps
Apply grease to litter grease points
Bed Serial No.
Completed By:_________________________________ Date:_____________