Maintenance Procedures Base
61
LIFT MOTOR ISOLATION PLATE REPLACEMENT (CONTINUED)
6. Remove the motor coupler (D) and bushings (E) through the access holes.
7. Using a 3/16 punch and hammer, remove the top roll pin on the manual drive hex shaft coupling and set
the shaft and coupler assembly aside.
8. Using a 3/8 open end wrench, remove the four standoffs (F) from the lift housing and lower the isolation
plate (B) with the attached standoffs.
9. Using a 3/8 socket, remove the two nuts and washers (A) and the two standoffs from one end of the
isolation plate and set the nuts, washers and standoffs aside.
10. Rotate the end of the isolation plate with the standoffs removed toward the access opening and tilt it
up and out of the opening.
11. Remove the two remaining standoffs from the isolation plate. Attach the two standoffs to one end of
the new isolation plate.
12. Reverse the above steps to reattach all components to the bed.
CAUTION
When reinstalling the standoffs to the lift housing, do not use a wrench to start the threads of the standoffs
and take care to unstall them straight or damage to the standoffs could occur.
NOTE
When reinstalling the lift motor, use the access holes in the lift housing to view the alignment of the motor
drive shaft to the motor coupler. The drive shaft on the motor probably will have to be turned to be aligned
with the coupler. Use a 7/16 open end wrench to turn the drive shaft of the motor.