Maintenance Procedures Siderail
95
HEAD END SIDERAIL CABLE REPLACEMENT
Required Tools:
#2 Phillips Screwdriver or T27 Torx Side Cutters
NOTE
Because of fastener changes, it will be necessary to use either a Phillips screwdriver or a T27 Torx to remove
the bolts securing the litter access cover.
Procedure:
1. Run the head section fully up.
2. Unplug the bed power cord from the wall socket.
3. Using a #2 Phillips screwdriver or a T27 Torx, remove the four screws securing the head end litter access
cover to the litter and remove the cover.
4. Using a #2 Phillips screwdriver, remove the eight screws holding the siderail cover and remove the cover.
5. Put the siderail in the down position.