Maintenance Procedures Base
64
LIFT MOTOR COUPLER REPLACEMENT
Required Tools:
#2 Phillips Screwdriver Bungee Cord (or equivalent) 5/16 Socket Wrench
5/16 Nut Driver 1/2 Socket (w/extension)
B
A
FOOT END BOTTOM VIEW
Procedure:
1. Unplug the bed power cord from the wall socket.
2. Using a #2 Phillips screwdriver, remove the three screws holding the upper lift cover to the base.
3. Using a 5/16 socket wrench, remove the six bolts holding the lower lift cover to the base and remove
the cover.
4. Using a 3/8 socket with extension, remove the four nuts and washers (A) holding the isolation plate (B)
to the lift housing and lower the lift motor and isolation plate assembly to allow access to the coupler.
Take care not to drop the isolation plate onto the base.