Maintenance Procedures Base
58
CASTER REPLACEMENT
Required Tools:
Floor Jack 1/2 Socket w/extension
#2 Phillips Screwdriver Bungee Cord (or equivalent) Wooden Brace
A
FOOT END BOTTOM VIEW
Replacement Procedure:
1. Unplug the power cord from the wall socket.
2. Using a #2 Phillips screwdriver, remove the three screws holding the upper lift cover to the base.
3. Be sure the beds brakes are on. Using a floor jack and wooden brace, lift the end of the bed with the
defective caster off the floor a few inches.
4. Using a 1/2 socket with extension under the base, remove the four castle nuts (A) holding the caster
assembly to the base.
5. Reverse the above procedure to install the replacement caster.
NOTE
The upper lift cover was raised to allow access to the four carriage bolts which need to be held when the four
castle nuts are reinstalled.