EasyManua.ls Logo

Stryker Secure 3000 - Head Motor Drive Isolator and CPR Decoupler Replacement

Stryker Secure 3000
203 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Maintenance Procedures Litter
79
HEAD MOTOR REMOVAL AND REPLACEMENT
6. Unplug the head motor cable (B) from the bed and capacitor.
WARNING
Support the head section with a wooden brace (2 x 4 or equivalent) before continuing or personal injury could
result.
7. Using a 5/16 nut driver, remove the two screws and washers (C) holding the motor mount to the litter.
8. Unplug the two cables (D) at the head motor capacitor.
9. To remove the head motor from the litter, push the motor toward the foot end and lift straight up on the
motor about 1. Pull the motor toward the head end and lift it straight up and out.
NOTE
The replacement motor will come with the attached CPR clutch and clutch bushings.
23. Using a 7/16 socket, remove the two bolts holding the motor to the mounting bracket.
NOTE
The mounting bracket will be reused with the replacement motor
24. Reverse the above steps to install the new motor assembly and reattach all litter panels. Be sure to attach
new cable ties wherever they were removed.
B
C
D
TOP VIEW