Loading...
Maintenance Procedures Litter
72
CAM AND CAM GUIDE REPLACEMENT HEAD AND KNEE MOTOR
Required Tools:
5/16 Nut Driver 7/16 Nut Driver #2 Phillips Screwdriver
7/16 Open End Wrench 3/8 Socket Wrench or T27 Torx
NOTE
Because of fastener changes, it will be necessary to use either a 3/8 socket wrench or a T27 Torx to remove
the bolts holding the ground straps to the motion interrupt pan and a Phillips screwdriver or a T27 Torx to
remove the bolts securing the litter access panels.
B
C
D
A
C
D
HEAD MOTOR
KNEE MOTOR
Procedure:
1. Unplug the power cord from the wall socket.
2. Using a 3/8 socket wrench or a T27 Torx, remove the two bolts holding the ground straps to the motion
interrupt pan.
3. Using a 5/16 nut driver, remove the two bolts holding the motion interrupt pan to the litter and remove
the pan.
4. Using a #2 Phillips screwdriver or a T27 Torx, remove the four screws holding the litter access cover to
the litter at the foot end of the bed and remove the cover.