Maintenance Procedures – Litter
78
HEAD MOTOR REMOVAL AND REPLACEMENT
Required Tools:
Side Cutters 5/16” Nut Driver 7/16” Socket Wrench
3/8” Socket Wrench #2 Phillips Screwdriver T27 Torx
NOTE
Because of fastener changes, it will be necessary to use either a Phillips screwdriver or a T27 Torx to remove
the bolts securing the litter access panel.
Procedure:
WARNING
The head motor is not grounded and may be an electrical hazard. Use caution when working around motors.
1. Run the head section up to approximately 30 – 35 for better access to the motor.
2. Unplug the bed power cord from the wall socket.
3. Using a #2 Phillips screwdriver or a T27 Torx, remove the four screws holding the head end litter access
cover to the litter and remove the cover.
4. Using side cutters, cut the cable tie holding the CPR release cable to the litter (to allow clearance for lifting
the motor out of the litter).
5. Using a 5/16” nut driver and a 3/8” socket wrench, remove the two screws (A) holding the CPR cable
bracket to the litter. Remove the cable from the bracket and remove the bracket.
A