Maintenance Procedures Litter
88
LOAD CELL REPLACEMENT
Required Tools:
#2 Phillips Screwdriver or T27 Torx 5/16 Nut Driver Side Cutters
1/4 Hex Allen Wrench
C
B
A
C
B
C
C
A
B
B
B
B
HEAD END
Procedure:
1. Unplug the bed power cord from the wall socket.
2. Using a #2 Phillips screwdriver or a T27 Torx, remove the four screws securing the litter access cover
at the end of the bed needing service and remove the cover.