EasyManua.ls Logo

Stryker Secure 3000 - Lift Motor Isolation Plate Replacement

Stryker Secure 3000
203 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Maintenance Procedures Base
60
LIFT MOTOR ISOLATION PLATE REPLACEMENT
Required Tools:
#2 Phillips Screwdriver 1/4 Nut Driver 3/8 Open End Wrench
3/8 Socket Wrench 3/16 Punch Hammer
NeedleNose Pliers 5/16 Nut Driver 7/16 Open End Wrench
3/16 Hex Allen Socket (w/6 ext.) 5/16 Socket Wrench
A
B
C
D
E
A
C
E
F
Procedure:
1. Unplug the bed power cord from the wall socket.
2. Using a #2 Phillips screwdriver, remove the three screws holding the upper lift cover to the base.
3. Using a 5/16 socket wrench, remove the six bolts securing the lower lift cover to the base and remove
the cover.
4. Using a 5/16 nut driver, remove the two screws holding the front and back metal access panels to the
lift housing assembly and remove the panels.
5. Using a 3/16 hex Allen socket with a 6 extension, remove the four bolts holding the lift motor to the base
and carefully lower the motor to the floor.

Table of Contents

Other manuals for Stryker Secure 3000

Related product manuals