Foot Board Operation Guide
25
WEIGH SYSTEM CONTROL PANEL GUIDE
1. LCD displays patient weight.
2. Push to zero bed.
3. Push when changing equipment on the bed.
4. Push to change weight from pounds to kilograms or back.
5. Push to turn weight system on.
6. Push to decrease numerical value of displayed weight.
7. Push to increase numerical value of displayed weight.
NOTE
After approximately 30 seconds of idle time, the scale display will turn off and will show the Trendelenberg
angle of the bed. Press SCALE ON to return to the weight display.
SYMBOL ACTION DISPLAY
To prepare bed for new patient:
Press and hold SCALE ON. LET GO FOR SCALE
Release the button after the WEIGHING...
display reads: XXX.X LB
LET GO FOR SCALE
Press and hold ZERO””HOLD TO ZERO WT.
RELEASE TO ZERO
Release ZERO””DO NOT TOUCH BED
0.0 LB