Foot Board Operation Guide
24
CENTER OF GRAVITY BED EXIT (OPTIONAL EQUIPMENT)
1. Push to activate Bed Exit function.
2. Push to deactivate Bed Exit function.
NOTE
If the scale system is in use, it will switch to off when Bed Exit is armed. Bed Exit will be temporarily disarmed
when the scale system is activated and will rearm when the scale system shuts off. When the bed is
equipped with scales, the scales must be properly zeroed for the Bed Exit System to function properly (see
page 27 for scale system usage instructions). If bed is not equipped with a scale system, follow the procedure
below.
1. Before putting a new patient on the bed: prepare bed for patient stay by adding linens and equipment to
the bed. Press and hold ARM and DISARM keys together for 5 seconds ARMED light will begin
to flash. Release ARM and DISARM keys and do not touch the bed until ARMED light stops flashing.
2. Once the new patient is on the bed: push ARM key and release (ARMED light will come on).
3. To deactivate Bed Exit, push DISARM. The ARMED and BED EXIT ON LEDs will turn off.
WARNING
The Bed Exit System is intended only to aid in the detection of a patient exiting the bed. It is NOT intended
to replace patient monitoring protocol. It signals when a patient is about to exit. Adding or subtracting objects
from the bed after arming the bed exit system may cause a reduction in the sensitivity of the bed exit system.