Maintenance Procedures – Litter
92
POWER CORD REPLACEMENT
Required Tools:
Heyco Pliers #2 Phillips Screwdriver or T27 Torx
Side Cutters 5/16” Nut Driver 1/4” Nut Driver
NOTE
Because of fastener changes, it will be necessary to use either a Phillips screwdriver or a T27 Torx to remove
the bolts securing the litter access panel.
Procedure:
1. Unplug the power cord from the wall receptacle.
2. Using a #2 Phillips screwdriver or a T27 Torx, remove the four screws securing the head end litter access
cover and remove the cover.
3. Using Heyco pliers, remove the two power cord strain relief connectors.
4. Using side cutters, cut all cable ties harnessing the power cord to other cables.
5. Unplug the power cord from the power board.
NOTE
Note the plug location for proper installation of the new power cord.
6. Using a 5/16” nut driver, remove the screw holding the power cord ground cable to the base.
NOTE
Note the arrangement of the star washers so the new ground wire will be installed properly.
7. Reverse the above steps to install the new power cord.
LITTER GREASE POINTS
1. All litter pivot joint bushings – “Syntech” synthetic grease
2. All Gatch and Fowler pivot link bushings – “Syntech” synthetic grease
3. Gatch drive screw and Gatch pan slide surface – “Syntech” synthetic grease
4. Fowler drive screw – “Mobil 28” grease