Maintenance Procedures Base
63
LIFT HOUSING REMOVAL AND REPLACEMENT (CONTINUED)
B
FOOT END BOTTOM VIEW
9. Lift the litter top up and support it about 6 above the lift screws with sawhorses or the equivalent.
10. Under the base, using a 9/16 socket, remove the four nuts (B) holding the lift housing to the base.
11. Lift up and out on the lift housing assembly to remove it from the base.
CAUTION
See pages 68 and 69 for proper reattachment procedure for power and sensor coil cords.
12. Reverse the above steps to reinstall the lift housing assembly after service is completed.
NOTE
The procedure for lift housing removal and replacement is the same for both ends of the bed.