Maintenance Procedures Base
65
LIFT MOTOR COUPLER REPLACEMENT (CONTINUED)
C
E
D
5. Using a 5/16 nut driver, remove the two screws (C) holding the metal access panel (D) to the lift housing.
6. The motor coupler (E) can now be removed from the lift housing through the access hole.
7. Reverse the above steps to install the new motor coupler and bushings.