Maintenance Procedures Base
56
STEER WHEEL CABLE REPLACEMENT AND ADJUSTMENT (CONTINUED)
E
6. Using two 3/16 hex socket drivers, loosen the bolt in the cable retainer (E) and slide the retainer off the
cable. Pull the cable out of the bed.
NOTE
If the cable sheathing is not damaged, pull the cable from the head end of the bed and remove it from the
sheathing, leaving the sheathing on the bed. Push the new cable through the sheathing from the head end.
7. Reverse the above steps to attach the new cable.