6. Troubleshooting
For repair and technical support, contact maintenance service center address
described in the back cover.
Arc problems
The electrodes typically need replacement after 1000 splices. Some common
symptoms that indicate the electrodes need replacing are:
٠Fluttering or unstable arc observed on the monitor
٠Sizzling noise while arcing
٠Bubbles in the fibers after splicing
٠Fiber burned in half
٠Diameter faults
٠High or inconsistent splice losses
Refer to page 35, “Replacing electrodes” for procedures.
If an electrode tip touches something, it will be deformed, causing poor arcing
problems. Take care of the handling of electrodes.
Fiber breaking
When the splicing process is complete, a proof test may be performed on the fibers
while in the fiber chucks. If the fibers are breaking when the proof test is performed,
re-do an arc test. If the arc power level is too weak, the splice may be poor, resulting in
fiber breaking.
If the fibers are breaking in spite of a good arc test result, clean the V-grooves and the
bare fiber pads completely. Deterioration of a jacket remover/fiber cleaver may lead to
fiber breaking. Clean the jacket remover/fiber cleaver completely.
Splicer does not power up
If the fusion splicer fails to turn on when the ON key is pressed, check the following:
٠Verify that the power supply module or battery module is installed in the module bay
properly
٠Verify that the power plug is seated properly (the power cord is connected to the
power supply module.)
٠Verify that the LED of the power supply module lights up.
٠If using battery operation, ensure that the battery module is fully charged.
If the splicer still does not power up after checking the above, contact our maintenance
service center.
50