7.3.4 Definition of calibration parameter
You can calibrate the electrode by one point or up to three points of standard solutions by
any sequence. As different calibration point method is applied, the definition of the zero point
and slope different.
Calibration point Determination The showed calibration value
One point
calibration
Asy
Zero point (Asy, offset or Zero point)= Asy
1.If not calibrated, Slope = Theoretical slope
2.If calibrated, Slope = Slope of last calibration
Above Two
points calibration
Asy
Slope
Zero point (Asy, offset or Zero point)= Asy
Slope = Slope
*
Note: To obtain a new zero point(Asy) and Slope
by applying linear regression.