3.6.3 Error Message
Message Alarm Cause Solution
RAM Fault The chip RAM is damaged. Replace the chip RAM.
EEPROM Fault The chip EEPROM is damaged. Replace the chip EEPROM.
Parameter Error The chip EEPROM is damaged. Replace the chip EEPROM.
Weighing Signal
Error
Weighing signal reversed or not
connected.
Connect the loadcell correctly.
ADC Fault The chip ADC is damaged. Replace the ADC module.
Over ADC Range
Weighing signal exceeds A/D
conversion range.
1. Check if the loadcell is connected.
2. Check if the capacity of loadcell is too small.
3. Check if the loading weight is too big.
Overload Alarm
Gross Weight > (Scale Capacity + 9
× Scale Division).
1 . Check if the set value of parameter [102]
Scale Capacity is reasonable.
2. Check if the loadcell is connected.
3. Check if the capacity of loadcell is too small.
4. Check if the loading weight is too big.
Internal Data-bus
Fault
Please contact manufacturer.