6.3.2 Calibration Parameters
No. Sign Range Default Description REG
120
Auto Zero
Tracking Rate
0.0~25.0 0.0
Auto Zero Tracking Rate
Set value = 0: No ‘Auto Zero Tracking’.
Set value > 0: Only when ‘Weight is stable’ and
‘Zero Variation
Rate < Auto Zero Tracking
Rate’, it’s allowed to enter Auto Zero Tracking
process.
Refer to p
arameter [108] ‘Stablity Judging
Range’ and [109] ‘Stablity Judging Time’。
40151
121
Auto Zero
Tracking Time
5~99 10 Auto Zero Tracking Time [s] 40153
122
Auto Zero
Tracking Range
0.1~50.0 5.0
Auto Zero Tracking Range [d: Division]
Only when the zero variation in ‘
Tracking Time’ is within this range, the result of
Auto Zero Tracking will be valid.
40155
123
Zero Fine
Adjusting Range
0~50000 50
Zero Fine Adjusting Range
Only when the weight variation caused by Zero
Valu
e changing is within this range, the
operation of ‘Zero Fine Adjustment’
valid.
Suggestion: Set value ≤ (Scale Capacity ×
4%).
40157
124
Range of
Displaying Zero
0~99 5
Range of Displaying Zero
When the absolute value of weight is within this
range, the real-time weight will return to zero.
40159
Please preset it to zero for metrological
certification in order that the minimum
division value can be displayed.
125
Initial Zero Adj.
after Power-on
0~1 0
Initial Zero Adjustment after Power-on
0: OFF
1: ON [without Power-down Protection]
40161
126
Initial Zero
Adjusting Time
0~1800 30 Initial Zero Adjusting Time [s] 40163
127
Initial Zero
Adjusting Range
0~50000 50
Initial Zero Adjusting Range
Suggestion: Set value ≤ (Scale Capacity ×
20%).
40165