42
6.4.2 Loss Calibration
Do Loss Caliration to correct Span Coefficient acccording to the weight of the materials dumped from
the weighing hopper.
Operation Steps:
Step0: Press the button#1/#2/#3/#4to let its status bar being red for selecting ‘Weigher
No.’.
Step1: Feed some materials into the weighing hopper, then stop the feeding process. After the real-time
weight being stable, pressFeedEndto get
AD0 Value.
Step2: Dump the partial or all materials in the weighing hopper into a container, then stop the dumping
process. After the real-time weight being stable, pressLossEndto get
AD1 Value.
Name Operation State Indicator
High Manual Start/Stop High-speed Feeding. Green: High-speed Feeding state.
Low Manual Start/Stop Low-speed Feeding. Green: Low-speed Feeding state.
Dump Manual Start/Stop Dumping. Green: Dumping state.