MPC-385 SERIES OPERATION MANUAL – REV. 3.21K (20201120)
6. MAINTENANCE
Routine cleaning of the MPC-385-series system is required to prevent excessive dust
accumulations. This is done by wiping all exterior surfaces with a dry, soft, cotton cloth.
All cables and connections should be periodically inspected to make sure that all connections
are made well and that all connectors are well and evenly seated.
CAUTION: The MP-285/M electromechanical is a precision-machined part, mounted on three
stepper motor shafts. As such, it DOES NOT REQUIRE LUBRICATION. Attempting to
lubricate any part of the electromechanical assembly will void the warranty and may harm
one or more of the motors.