EasyManua.ls Logo

Sutter Instrument MPC-200 - ROE-200 Controls Overview

Sutter Instrument MPC-200
73 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
MPC-385 SERIES OPERATION MANUAL REV. 3.21K (20201120)
15
If you toggle from Drive A to Drive B and back again you will see that the display coordinates
and Mode settings are maintained for Drive A while you are using Drive B and vice versa.
The remaining functions of the ROE are explained in the next section.
2.3 Main Controls on the ROE-200
2.3.1 White Buttons:
DIAG/NORM: Pressing the DIAG/NORM button will cause the green LED near the button to
light, indicating the MPC-200/ROE-200 is in Diagonal mode. In this mode, rotation of the Z-
axis knob produces diagonal movement. A second press will put the manipulator back into
Normal mode. When in diagonal mode, the X and Y knobs remain active, allowing you to
readjust the X and Y positioning of the pipette as you approach a cell in diagonal mode. Angle
of diagonal mode movement is set via DIP switches on the back of the MPC-200 controller
(see controller DIP switch setting instructions on Pages 19-21). When using MODE 9 (MODE
toggle set to 9), Diagonal mode produces short, quick, impulse-like movement that may be
useful in sharp pipette impalements.
When you switch to Diagonal mode, the ROE-200 display is changed from absolute to relative
coordinates and the current location is set to 0,0,0. This allows users to invoke relative
measurements using the display as a measuring device. A fourth coordinate that gives
movement along the diagonal is also added for users who wish to measure the movement of
along the axis coaxial with a pipette. When you return to Normal mode, the absolute
coordinate system is recovered. The relative coordinate feature can be disabled via DIP
switch 2 on back of the ROE-200.
Figure 2-2. Top view of ROE-200.
HOME: When pressed, the manipulator will make a move along a stereotypic path to the
location 0,0,0 or "home”. Home is the location where you would most likely exchange your
pipette and is maximal up on the Z-axis, maximal right on the X-axis (maximal left on a left-
handed manipulator) and maximal front on the Y-axis. The stereotypic path of the movement

Table of Contents

Related product manuals