SV 104IS USER’S MANUAL ______________________________________________________
28
Notice: To charge a fully discharged battery, it would take approximately 4 hours.
Notice: To charge multiple dosimeters at the same time via the docking station, ensure
your 12V AC/DC power supply such as the SA 33 is capable in terms of providing enough current
performance.
Current estimation: Assume the current consumption on 0.24A level, and calculate required minimum power
supply current (5 x 0.24A = 1.2A).
Power wattage estimation: With nominal voltage of 12V the power supply wattage should be minimum
2.88W per dosimeter (5 x 2.88W = 14.4W).
Notice: Charging is allowed only in safe area. See the SPECIAL PRECAUTIONS WHEN
USING AND CHARGING LITHIUM BATTERIES clause on page 4.
4.2 Before you turn the instrument on
There are only a few things to remember:
Make sure the microphone is properly fastened onto mounting head before turning on the
instrument.
When in use for a measurement run, always use the SA 122IS windshield provided.
Notice: On new products there is a display shipping protection foil which is used on new
products being sent to protect from accidental scratches. It is ~25x15mm foil rectangle.
Peel off the protection foil and throw away.