SV 104IS USER’S MANUAL ______________________________________________________
72
6.9 Troubleshooting
1. Upon connection onto the docking station port, if automatic charging is not started: check the
docking station LED colour for state of charging (chapter 4.1).
2. If the incorrect time or date is displayed when turning on the instrument connect the device to the
computer and use Supervisor software to set the time and date (see) ensuring PC clock is set
correctly.
3. In case the instrument is not able to turn on ensure the unit is charged by connecting onto the
docking station. This ensures the battery is not exhausted. Then proceed with hardware reset
(chapter 4.15)
4. In case your dosimeter does not respond proceed with turn-off/turn-on procedure (chapter 4.3),
and hardware reset of the instrument (chapter 4.15).
5. In case the measurement of the sound level is frozen or set to a fixed value proceed with turn-
off/turn-on procedure (chapter 4.3), then with hardware reset of the instrument (chapter 4.15).
6. In case the reset does not help proceed to chapter 7.