EasyManua.ls Logo

Tektronix 222 - Page 140

Tektronix 222
167 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Checking Performance
-222
DSO Operators
3. Check Vertical POSITION Range
a. Set the CH
1
VOLTSIDIV setting to 0.2 V.
b. Set the Calibration Generator to output a 5-V, standard-
amplitude signal.
c. Adjust the Vertical POSITION control fully clockwise.
d. CHECK-for the bottom of the signal to be no more than 2
divisions below the center horizontal graticule line.
e. Adjust the Vertical POSITION control fully counterclockwise.
f.
CHECK-for the top of the signal to be no more than
2
divisions above the center horizontal graticule line.
g. SET: CH 1 Coupling CHI
OFF
CH 2 Coupling AC
CH 2 VOLTSIDIV 0.2 V
h. Disconnect the Channel 1 probe tip from the test equipment
and connect the Channel 2 probe tip.
i. Repeat parts c, d, e, and f for CH 2.
4.
Check Variable VOLTSIDIV Range
a. SET: CH 2 Coupling DC
CH 2 VOLTSIDIV 50 mV
b. Set the Calibration Generator for a standard-amplitude output
of 0.2 V.
c. Position the bottom of the signal on the center vertical graticule
line.
d. Press in and rotate the Vertical POSITION knob
WAR VOLTS/
DIV control when held in) counterclockwise until the display ampli-
tude stops decreasing.
e. CHECK-for less than
1.6
division of signal amplitude.

Related product manuals