Performance Tests (MSO70000/C Series, DSA/DPO70000B/C Series, and DPO7000 Series)
Table 3-7: Anal
og bandwidth, < 3.5 GHz models
Test frequency
Vertical scale
Reference
amplitude
Horizontal
scale DPO7254 DPO7104 DPO7054 -3 dB
5mV 30mV 1ns
2.0 GHz 1 G Hz
500 MHz ≥21.2 mV
10 mV 60 mV 1 ns
2.501 GHz 1 GHz
500 MHz ≥42.4 mV
20 mV 120 mV 1 ns
2.501 GHz 1 GHz
500 MHz ≥84.8 mV
50 mV 300 mV 1 ns
2.501 GHz 1 GHz
500 MHz ≥212 mV
100 mV 600 mV 1 ns
2.501 GHz 1 GHz
500 MHz ≥424 mV
200 mV 1.2 V 1 ns
2.501 GHz 1 GHz
500 MHz ≥848 mV
500 mV 3 V
1
1ns
2.501 GHz 1 GHz
500 MHz ≥2.12 V
1
1V 5V
1
1ns
2.501 GHz 1 GHz
500 MHz ≥3.535 V
1
1
If your generator cannot output the required amplitude, determine its maximum output a t the Test frequency, and use this for the reference amplitude. The -3 dB
limit c
an be calculated as: 0.707 × reference amplitude.
c. Set the vertical scale:
For the channel you are testing, set the vertical Scale to the next setting in
the table, starting with the 100 mV setting. (See Table 3-7.)
d. Set the triggering coupling: From the button bar, touch Trigger, then
select Coupling DC.
e. Display the test signal: Do the following subparts to first display the
reference signal and then the test signal.
From the button bar touch Measure; then select the Time tab.
Touch the Freq button to measure the frequency of the current channel.
Select the Ampl tab. Touch the Amplitude button.
Touch the Statistics button, then touch the Mean button.
T
ouch the X (Close) button.
Set the generator output so the Chx Amplitude mean readout equals
the reference amplitude in the table. This corresponds to the vertical
scale set in substep c. (See Table 3-7.)
Record the reading on the power meter.
Set the trigger as follows:
Press the front-panel Push-Set 50% as necessary to trigger a stable
display. At full bandwidth, you may also want to make small, manual
adjustments to the trigger level. You can use the Trigger Level knob to
do this.
3–62 MSO70000/C, DSA70000B/C, DPO7000B/C, DPO7000, MSO5000, DPO5000 Series