9 Performing the measurement
48
9.6.2
Flue Gas Analysis
To maintain the measuring accuracy of the instrument, the correct fuel
must be selected or configured.
Tap (Fuels)
> Select fuel.
To achieve usable measurement results, the test time of a flue gas
measurement should be at least 3 minutes and the measuring
instrument should display stable readings.
If a separate CO Air free measurement has not yet been performed,
this value is calculated using the readings from the flue gas probe and
continuously updated.
1
Call up function: | Flue Gas Analysis
2
Start measurement: Tap .
Zeroing takes place.
Readings are displayed.
If the Draft
measurement parameter is enabled in the reading display, a
draft measurement is automatically started in parallel to the flue gas
measurement. In the List measurement data view, the parallel draft
measurement can be stopped/restarted. This draft measurement is
performed separately to a measurement of the Draft measurement
type.