10 Maintenance
69
Clean modular flue gas probe
Disconnect flue gas probe from the measuring instrument.
Release probe catch by pressing the key on the probe handle and
remove probe module.
Blow compressed air through flue gas
ducts of the probe module and probe
handle (see illustration).
Do not use a
brush!
Fit probe module onto the probe handle and click into place.
Replace the probe module
Disconnect the flue gas probe from the measuring instrument.
Press the key on the top of the probe
handle and remove the probe module.
Fit new probe module and click into place.
Check/replace particle filter
Check particle filter
• Particle filters of the modular flue gas probe must be checked regularly for
contamination: check visually by looking through the window of the filter
chamber.
• If there is visible contamination or inadequate pump flow, replace the filter.