9 Performing the measurement
52
O
2
air (EU regulation)
2
dual wall clearance probe (0632 1260) must be connected.
Call up function: | O
2
air
Start measurement: tap .
Reading is displayed.
End measurement: tap .
Clock Meter
The function is only available if the chosen fuel is a gas.
The gas burner capacity is calculated by means of the gas amount consumed.
To this end, a gas amount is input, and its consumption read out at the gas
meter.
Call up function: | Clock Meter
Set the gas amount to watch for at the gas meter.
Set the heating value of the burnt gas.
Start measurement: tap .
The test time is displayed.
When the set gas amount is reached: tap .
The calculated gas flow and the gas burner capacity (in BTU/h) are
displayed.
Edit values
All values that can be modified have a dotted underlining.