10 Maintenance
70
Replace particle filter
The filter chamber may contain condensate. This is not a malfunction
and will not cause incorrect readings.
Open filter chamber: turn slightly
counter-clockwise.
Remove filter chamber.
Remove filter plate and replace it
with a new one (0554 3385).
Attach the filter chamber and lock it:
turn slightly clockwise.