© 2020 Thorlabs GmbH
4 Write Your Own Application
45
OUTPut subsystem commands
Enables (ON) or disables (OFF) LED output
Select output terminal: 1 = 10A/12Pin connector
(LED1); 2 = 2A/4Pin connector (LED2);
Returns selected output terminal
Abort any running LED head presence test
Testing presence and properties of LEDs
Initiate presence test procedure on the specified ter-
minal
Query status of the presence test procedure as an
<NF1> value:
0 = The presence test procedure is running
1 = The latest presence test detected no LED
2 = The latest presence test detected a custom LED
without onboard head info memory
3 = The latest presence test detected a LED head
with onboard head info memory
Returns interlock circuit protection tripped (1) or un-
tripped (0)
Returns driver (console) over temperature protection
tripped tripped (1) or untripped (0)
Returns head over temperature protection tripped
tripped (1) or untripped (0)
SENSe3 LED current sensing subsystem commands
Path to LED current sensing
Returns the measured LED current. Parameters MIN
and MAX return the measurement value range pos-
sible in the currently active driver configuration (de-
pends on the compliance voltage required by the
LED head)
SENSe4 LED voltage sensing subsystem commands
Path to LED voltage sensing
Returns the measured LED voltage. Parameters MIN
and MAX return the measurement value range pos-
sible.
SENSe5 LED temperature sensing subsystem commands
Path to LED temperature sensing
[:TEMPerature]? [{MIN|MAX}]
Returns the measured LED temperature. Parameters
MIN and MAX return the measurement value range
possible.