TC300 Heater Temperature Controller Chapter 8: Command Line Operation
Page 32 CTN017856-D02
Set target temperature min value of Channel 1, in unit of 1°C.
TMIN1?
Query target temperature min value of Channel 1, in unit of 1°C.
Return:
n
TMIN2=n
Set target temperature min value of Channel 2, in unit of 1°C.
TMIN2?
Query target temperature min value of Channel 2, in unit of 1°C.
Return:
n
8.3.8. VOLT1?; VOLT2? – Actual Output Voltage of TC300
VOLT command is used to query the actual output voltage of the TC300 device, the range is from 0.1 to 24.0 V.
VOLT1?
Query actual output voltage of Channel 1, in unit of V, resolution is 0.1V
VOLT2?
Query actual output voltage of Channel 1, in unit of V, resolution is 0.1V
8.3.9. CURR1=n; CURR1?; CURR2?; CURR2=n – Output Current of TC300
CURR command is used to set and query the output current of the TC300 device, the range is from -2000 to
+2000 mA.
CURR1=n
Set output current of Channel 1, in unit of 1 mA.
CURR1?
Query output current of Channel 1, in unit of 1 mA.
Return:
n
CURR2=n
Set output current of Channel 2, in unit of 1 mA.
CURR2?
Query output current of Channel 2, in unit of 1 mA.
Return:
n
8.3.10. CMAX1=n; CMAX1?; CMAX2=n; CMAX2? – Output Current Max Value of TC300
CMAX command is used to set and query the output current max value of the TC300 device, the range is from
0.010 to 2.000 A.
CMAX1=n
Set the CH1 output current max value, n ranges from 10 to 2000 mA, resolution is 1 mA.
CMAX1?