TC300 Heater Temperature Controller Chapter 8: Command Line Operation
Rev B, January 5, 2022 Page 33
Get the CH1 output current max value in the unit of A (range:0.010 to 2.000 A, resolution is 0.001 A).
Return:
n/1000
CMAX2=n
Set the CH2 output current max value, n ranges from 10 to 2000 mA, resolution is 1 mA.
CMAX2?
Get the CH2 output current max value in the unit of A (range:0.010 to 2.000 A, resolution is 0.001 A).
Return:
n/1000
8.3.11. VMAX1=n; VMAX1?; VMAX2=n; VMAX2? – Output Voltage Max value of TC300
VMAX command is used to set and query the output voltage max value of the TC300 device, the range is from 0.1
to 24.0 V.
VMAX1=n
Set the maximum allowed voltage of Channel 1 to n/10 V, n ranges from 1 to 240, corresponding to 0.1 to 24.0 V.
VMAX1?
Query the maximum allowed voltage of Channel 1, in unit of V, resolution is 0.1 V.
Return:
n/10
VMAX2=n
Set the maximum allowed voltage of Channel 2 to n/10 V, n ranges from 1 to 240, corresponding to 0.1 to 24.0 V.
VMAX2?
Query the maximum allowed voltage of Channel 2, in unit of V, resolution is 0.1 V.
Return:
n/10
8.3.12. CSET1?; CSET2? – Output Current Set Value of TC300
CSET command is used to query the output current set value of the TC300 device, the range is from -2000 to
+2000 mA.
CSET1?
Query output current set value of Channel 1, in unit of 1 mA.
CSET2?
Query output current set value of Channel 2, in unit of 1 mA.
8.3.13. MOD 1 =n; MOD 1?; MOD =n; MOD 2? – Test Mode of TC300
MOD command is used to set and query the test mode of TC300 device.
MOD1=n
Set test mode of Channel 1
n=0: Heater
n=1: TEC