TC300 Heater Temperature Controller Chapter 8: Command Line Operation
Page 34 CTN017856-D02
n=1: Constant Current
MOD1?
Query test mode of Channel 1
Return:
n
MOD2=n
Set test mode of Channel 2
n=0: Heater
n=1: TEC
n=2: Constant Current
MOD2?
Query test mode of Channel 2
Return:
n
8.3.14. OFFSET1=n; OFFSET1?; OFFSET2=n; OFFSET2? – Sensor Offset Value of TC300
OFFSET command is used to set and query the sensor offset value of the TC300 device, the range is from -10.0
to +10.0 °C.
OFFSET1=n
Set the sensor offset value of Channel 1 to n/10V, n ranges from -100 to +100, corresponding to -10 to +10 °C
OFFSET1?
Query the sensor offset value of Channel 1 in °C.
Return:
n/10
OFFSET2=n
Set the sensor offset value of Channel 2 to n/10 V, n ranges from -100 to +100, corresponding to -10 to +10 °C
OFFSET2?
Query the sensor offset value of Channel 2 in °C.
Return:
n/10
8.3.15. TYPE1=n; TYPE1?; TYPE2=n; TYPE2? – Sensor Type of TC300
TYPE command is used to set and query the sensor type of the TC300 device. Channel 1 and Channel 2 share the
EXT1/EXT2. Channel 2 cannnot select EXT1/EXT2 when it is configured for Channel 1 or the other way.
TYPE1=n
Set sensor type of Channel 1
n=0: PT100
n=1: PT1000
n=2: NTC1
n=3: NTC2
n=4: Thermo C.
n=5: AD590