TC300 Heater Temperature Controller Chapter 8: Command Line Operation
Page 38 CTN017856-D02
Set Steinhart-Hart C coefficient of Channel 1 to n/10000, the range of n is -99999 to 99999 with a resolution of 1.
HARTC1?
Query Steinhart-Hart C coefficient of Channel 1, Range: 0 to 9.9999, Resolution: 0.0001.
Return:
n/10000
HARTA2=n
Set Steinhart-Hart A coefficient of Channel 2 to n/10000, the range of n is -99999 to 99999 with a resolution of 1.
HARTA2?
Query Steinhart-Hart A coefficient of Channel 2, Range: 0 to 9.9999, Resolution: 0.0001.
Return:
n/10000
HARTB2=n
Set Steinhart-Hart B coefficient of Channel 2 to n/10000, the range of n is -99999 to 99999 with a resolution of 1.
HARTB2?
Query Steinhart-Hart B coefficient of Channel 2, Range: 0 to 9.9999, Resolution: 0.0001Return:
n/10000
HARTC2=n
Set Steinhart-Hart C coefficient of Channel 2 to n/10000, the range of n is -99999 to 99999 with a resolution of 1.
HARTC2?
Query Steinhart-Hart C coefficient of Channel 2, Range: 0 to 9.9999, Resolution: 0.0001.
Return:
n/10000
8.3.22. TRIG1=n ; TRIG1? TRIG2=n ; TRIG2?– Trigger Mode of 300
TRIG command is used to set and query trigger mode of the TC300 device.
TRIG1=n
Set trigger mode of Channel 1, when n=1, Output, when n=0, Input
TRIG1?
Query trigger mode of Channel 1, when n=1, Output, when n=0, Input
Return:
n
TRIG2=n
Enable trigger mode of Channel 2, when n=1, Output, when n=0, Input
TRIG2?
Query trigger mode of Channel 2, when n=1, Output, when n=0, Input
Return:
n
8.3.23. BRIGHT=n; BRIGHT? – Brightness of TC300
BRIGHT command is used to set and query brightness of TC300 Display, the range is from 10 to 100%.
BRIGHT=n
Set brightness of TC300 Display in %. The range is from 10 to 100%. Resolution is 1 %.
BRIGHT?
Get brightness of TC300 Display in %.