94 WFS Series
NOTE
In Highspeed Mode there is an enhanced risk of mismeasurement because not all
violations of save measurement conditions can be detected. For instance, the image
saturation level remains unknown.
Spot centroids mismatched
There is a violation of the spot centroid positions detected within the camera with its
calculation windows defined when Highspeed Mode was entered. That is, the spot
shifts went too far and cannot be detected accurately anymore.
Switch off and on Highspeed Mode with option 'Adapt centroids' enabled to
overcome this situation.